Designed for testing filter systems ranging from small glove boxes to large cleanroom plenums, ATI Laskin Nozzle generators provide the ideal solution for creating a polydispersed, submicron aerosol to certify and qualify in-situ filter systems.
Aerosol concentration output varies by generator model; however, concentrations can be achieved from 100 μg/l down to 10 μg/l at airflows ranging from 405 to 8,100 CFM.
PERFORMANCE SPECIFICATIONS |
Generations |
Thermal condensations |
Warm Up Times |
<90 second to reached operation temperatures of 360º C (699º F) |
Systems Flowing Rates |
829-120,000 m3/hr. (500 – 70,000 cfm) |
Maximum Oil Consumptions |
19.9 g/min (0.781 oz.) |
Running times |
40 minute (at oil consumptions) |
Particles Distributions |
Meet ANSI/ASME N509/510 |
Dimension (L x W x H) |
42cm x 12cm x 34cm (17” x5”x15”) |
Weights |
approximately 8 kg empty (17.6 Lbs.) |
Standard |
ISO14644-3, CE RoHS 2 (2011/65/EU directive) |
PRODUCT SPECIFICATIONS |
Dimension |
42 cm x 12 cm x 36 cm (17″ x 5″ x 15″) |
Weights |
9 kg (20 lb) empty;
17 kg (40 lb) |
Power Consumptions |
1040 Watt |
Liquid Aerosol Reagent Capacity |
1.2 Liter (65 fl oz) |
Compress Insert Gas Inputs |
1.70 Bar ± 0.08 Bar (26 psi ± 1.0 psi) |
Compress Insert Gas Consumptions Rates |
6.5 L/min (0.22 cfm) |
Duct Systems Requirements |
Negative pressure required. Positive duct pressure requires ATI Positive Injection Pump (sold separately). |
Duty Cycles |
Intermittent ON/OFF or <4 hours ON and >1 hour off |